Influence of electrode placement on effective field strength in the superficial digital flexor tendon of horses
2006
Lin, Y.L. | Moolenar, H. | Weeren, P.R van | Lest, C.H.A van de
Objective-To determine the relationship between the output of an electrical treatment device and the effective field strength in the superficial digital flexor tendon of horses. Sample Population-Cadaver horse forelimbs without visible defects (n = 8) and 1 live pony. Procedure-Microcurrents were generated by a microcurrent electrical therapy device and applied in proximodistal, dorsopalmar, and mediolateral directions in the entire forelimbs, dissected tendons, and the pony with various output settings. Corresponding field strengths in the tendons were measured. Results-A linear relationship was detected between current and field strength in all conditions and in all 3 directions. In dissected tendons, significant differences were detected among all 3 directions, with highest field strength in the proximodistal direction and lowest in the dorsopalmar direction. In the entire forelimbs, field strength in the proximodistal direction was significantly lower than in the mediolateral direction. Results in the pony were similar to those in the entire forelimbs. Conclusions and Clinical Relevance-Electrode placement significantly affected field strength in the target tissue. Many surrounding structures caused considerable reduction of field strength in the target tissue. These factors should be taken into account when establishing protocols for electrical current-based therapeutic devices if these devices are proven clinically effective.
اظهر المزيد [+] اقل [-]الكلمات المفتاحية الخاصة بالمكنز الزراعي (أجروفوك)
المعلومات البيبليوغرافية
تم تزويد هذا السجل من قبل National Agricultural Library