Evaluation of three applanation tonometers in dogs
1990
Priehs, D.R. | Gum, G.G. | Whitley, R.D. | Moore, L.E.
The Mackay-Marg, Tono-Pen, and Challenger applanation tonometers were evaluated in vivo in 12 clinically normal eyes of 6 dogs. Tonometric measures of intraocular pressure (IOP) were compared with closed manometric IOP measurements from the anterior chamber of anesthetized dogs. The tonometers were evaluated at IOP that ranged from 5 to 100 mm of Hg. The Mackay-Marg tonometer was the most reliable instrument when evaluated at IOP from 5 to 100 mm of Hg (r2 = 0.996) and from 10 to 30 mm of Hg (r2 = 0.962). The Tono-Pen tonometer was also reliable (r2 = 0.967) over the range of IOP, but consistently overestimated IOP at lower pressures and underestimated IOP at higher pressures. The Mackay-Marg and Tono-Pen measurements were essentially linear. When evaluated from 10 to 30 mm of Hg, r2 was 0.828 for the Tono-Pen tonometer. The Challenger tonometer, although reliable over the full range of IOP (r2 = 0.965), proved to be less accurate, as indicated by lack of a good linear equation.