A human calorimeter for the direct and indirect measurement of 24 h energy expenditure
1978
Dauncey, M.J. | Murgatroyd, P.R. | Cole, T.J.
Extract: A calorimeter for the continuous measurement of heat production and heat loss in the human subject, for at least 24 h, is described. Sensible heat loss was measured using a water cooled heat exchanger, and the temperature of water entering the heat exchanger was controlled to maintain a mean temperature gradient of zero across the chamber walls. Evaporative heat loss was determined from ingoing and outgoing wet and dry bulb temperatures and air flow rates. Problems associated with the calculation of evaporative heat loss and the estimation of the volume of incoming air in open circuit systems are considered. The calibration, limits of accuracy, sources of error and experiments with subjects are discussed.