Direct Detection of a Sulfonate Ester Genotoxic Impurity by Atmospheric-Pressure Thermal Desorption–Extractive Electrospray–Mass Spectrometry
2013
Devenport, Neil A. | Sealey, Laura C. | Alruways, Faisal H. | Weston, Daniel J. | Reynolds, James C. | Creaser, Colin S.
A direct, ambient ionization method has been developed using atmospheric pressure thermal desorption–extractive electrospray–mass spectrometry (AP/TD-EESI-MS) for the detection of the genotoxic impurity (GTI) methyl p-toluenesulfonate (MTS) in a surrogate pharmaceutical matrix. A custom-made thermal desorption probe was used to the desorb and vaporize MTS from the solid state, by rapid heating to 200 °C then cooling to ambient temperature, with a cycle time of 6 min. The detection of MTS using EESI with a sodium acetate doped solvent to generate the [MTS+Na]⁺ adduct ion provided a significant sensitivity enhancement relative to the [M+H]⁺ ion generated using a 0.1% formic acid solvent modifier. The MTS detection limit is over an order of magnitude below the long-term daily threshold of toxicological concern (TTC) of 1.5 μg/g and the potential for quantitative analysis has been determined using starch as a surrogate active pharmaceutical ingredient (API).
Show more [+] Less [-]AGROVOC Keywords
Bibliographic information