A transducer for measuring force on surgical sutures
2010
The objective of this study was to validate, both in vitro and in an ex vivo model, a technique for the measurement of forces exerted on surgical sutures. For this purpose, a stainless steel E-type buckle force transducer was designed and constructed. A strain gauge was mounted on the central beam of the transducer to measure transducer deformation. The transducer was tested and calibrated on a single strand of surgical suture during cyclic loading. Further validation was performed using a previously published cadaveric model of laryngoplasty in the horse. Linear regression of transducer output with actual force during calibration tests resulted in mean R2 values of 1.00, 0.99, and 0.99 for rising slope, falling slope, and overall slope, respectively. The R2 was not less than 0.96 across an average of 75 cycles per test. The difference between rising slope and falling slope was 4%. Over 45 846 samples, the predicted force from transducer output showed a mean error of 4%. In vitro validation produced an adjusted R2 of 0.99 when the force on the suture was regressed against translaryngeal pressure in a mixed-effects model. E-type buckle force transducers showed a highly linear output over a physiological force range when applied to surgical suture in vitro and in an ex vivo model of laryngoplasty. With appropriate calibration and short-term in vivo implantation, these transducers may advance our knowledge of the mechanisms of success and failure of techniques, such as laryngoplasty, that use structural suture implants.
Show more [+] Less [-]AGROVOC Keywords
Bibliographic information
This bibliographic record has been provided by National Agricultural Library