Assessment of a commercially available veterinary blood pressure device used on awake and anesthetized dogs
2019
Cremer, Jeannette | Da Cunha, Anderson F. | Paul, Linda J. | Liu, Qinqi | Acierno, Marc J.
OBJECTIVE To compare results of a commercially available device for oscillometrically measured blood pressure (OBP) with invasively measured blood pressure (IBP) in awake and anesthetized dogs. ANIMALS 19 adult dogs (mean ± SD body weight, 17.8 ± 7.5 kg). PROCEDURES Blood pressures were measured in dogs while they were awake and anesthetized with isoflurane. The OBP was recorded on a thoracic limb, and IBP was simultaneously recorded from the median caudal artery. Agreement between OBP and IBP was evaluated with the Bland-Altman method. Guidelines of the American College of Veterinary Internal Medicine (ACVIM) were used for validation of the oscillometric device. RESULTS In awake dogs, mean bias of the oscillometric device was −11.12 mm Hg (95% limits of agreement [LOA], −61.14 to 38.90 mm Hg) for systolic arterial blood pressure (SAP), 9.39 mm Hg (LOA, −28.26 to 47.04 mm Hg) for diastolic arterial blood pressure (DAP), and −0.85 mm Hg (LOA, −40.54 to 38.84 mm Hg) for mean arterial blood pressure (MAP). In anesthetized dogs, mean bias was −12.27 mm Hg (LOA, −47.36 to 22.82 mm Hg) for SAP, −3.92 mm Hg (LOA, −25.28 to 17.44 mm Hg) for DAP, and −7.89 mm Hg (LOA, −32.31 to 16.53 mm Hg) for MAP. The oscillometric device did not fulfill ACVIM guidelines for the validation of such devices. CONCLUSIONS AND CLINICAL RELEVANCE Agreement between OBP and IBP results for awake and anesthetized dogs was poor. The oscillometric blood pressure device did not fulfill ACVIM guidelines for validation. Therefore, clinical use of this device cannot be recommended.
Show more [+] Less [-]AGROVOC Keywords
Bibliographic information
This bibliographic record has been provided by National Agricultural Library