Spontaneous alteration of blood pH by a bicarbonate buffer system during experimental hypercalcaemia in cows
2021
Ro, Younghye | Choi, Woojae | Hong, Leegon | Kim, Eunkyung | Choe, Eunhui | Kim, Danil
Maintaining mineral homeostasis as well as the secretion and metabolism of mineralotropic hormones is important for healthy of periparturient dairy cows. To increase the activity of mineralotropic hormones, blood pH can be adjusted. The purpose of this study was to investigate changes in blood pH and the mechanism of action of this change in induced hypercalcaemic cows. Six non-lactating Holstein cows were used in a 2 × 2 crossover design. To induce hypercalcaemia, calcium borogluconate was administered subcutaneously to experimental cows and normal saline was administered subcutaneously to control cows. Blood and urine samples were collected serially after administration. Whole blood without any anticoagulant was processed with a portable blood gas analyser. Plasma concentration and urinary excretion of calcium were measured. In hypercalcaemic cows, both blood and urine calcium levels were significantly increased at 8 h compared to those at 0 h (P < 0.05), and a spontaneous increase in blood pH was also observed. The calcium concentration in plasma was highest at 2 h after administration (3.02 ± 0.27 mmol/L). The change in pH correlated with that in bicarbonate (r = 0.781, P < 0.001) rather than that in partial pressure of CO₂ (r = 0.085, P = 0.424). Hypercalcaemia induced a spontaneous change in blood pH through the bicarbonate buffer system and this system may be a maintainer of calcium homeostasis.
Afficher plus [+] Moins [-]Evaluation of two benchtop blood gas analyzers for measurement of electrolyte concentrations in venous blood samples from dogs
2021
OBJECTIVE To assess agreement between 2 benchtop blood gas analyzers developed by 1 manufacturer (BGA 1 and BGA 2 [a newer model with reduced maintenance requirements]) and a reference chemistry analyzer for measurement of electrolyte (sodium, chloride, and potassium) in blood samples from dogs. ANIMALS 17 healthy staff- and student-owned dogs and 23 client-owned dogs admitted to an emergency and intensive care service. PROCEDURES Blood collected by venipuncture was placed in lithium heparin–containing tubes. Aliquots were analyzed immediately with each BGA. Samples were centrifuged, and plasma was analyzed with the reference analyzer. Results for each BGA were compared with results for the reference analyzer by Passing-Bablok regression analysis. Percentage differences between BGA and reference analyzer results were compared with published guidelines for total allowable error. RESULTS Proportional bias was detected for measurement of chloride concentration (slope, 0.7; 95% CI, 0.7 to 0.8), and constant positive bias was detected for measurement of chloride (y-intercept, 34, mmol/L; 95% CI, 16.9 to 38 mmol/L) and potassium (y-intercept, 0.1 mmol/L; 95% CI, 0.1 to 0.2 mmol/L) concentrations with BGA 1. There was no significant bias for measurement of potassium or chloride concentration with BGA 2 or sodium concentration with either BGA. Differences from the reference analyzer result exceeded total allowable error guidelines for ≥ 1 sample/analyte/BGA, but median observed measurement differences between each BGA and the reference analyzer did not. CONCLUSIONS AND CLINICAL RELEVANCE Good agreement with reference analyzer results was found for measurement of the selected electrolyte concentrations in canine blood samples with each BGA.
Afficher plus [+] Moins [-]