Oscillometric and Doppler arterial blood pressure measurement in conscious goats
2018
Szaluś-Jordanow, O. | Czopowicz, M. | Świerk, A. | Szpinda, O. | Garncarz, M. | Mickiewicz, M. | Moroz, A. | Bagnicka, E. | Kaba, J.
The objective of this study was to characterize arterial blood pressure (BP) measurements obtained by using 2 indirect methods, oscillometry and Doppler ultrasonic sphygmomanometry, in conscious goats. Agreement between systolic BP yielded by these 2 methods was then assessed. Sixty female dairy goats aged from 1.5 to 11.8 y (median: 5.5 y) were examined in a standing position with a cuff placed on the tail. All goats had a severe arthritic form of caprine arthritis-encephalitis. Three to 5 repeated measurements of each BP type were averaged for each goat and considered as a final measurement. With oscillometry, systolic blood pressure (O-SBP), diastolic blood pressure, and mean blood pressure, as well as heart rate (HR) were measured, while only systolic blood pressure was measured with Doppler (D-SBP). The O-SBP did not correlate with D-SBP [correlation coefficient (r) = 0.24, P = 0.067]; the mean difference (± standard deviation) was 24.5 ± 26.3 mmHg and limits of agreement were from -27.2 mmHg [95% confidence interval (CI): -39.0, -15.4 mmHg] to 76.1 mmHg (95% CI: 64.3, 87.9 mmHg). No significant linear correlation was found between any BPs and HR (r: -0.10 to 0.22) or age (r: -0.26 to 0.07) of the goats. The study showed that, while BP could be measured in conscious goats using both oscillometry and Doppler ultrasonic sphygmomanometry, the results obtained were so inconsistent that these methods could not be used interchangeably.