Alternative method for intramuscular fat analysis using common laboratory equipment
2015
Segura, J. | Calvo, Luis | Óvilo Martín, Cristina | González De Bulnes, Antonio | Olivares, A. | Cambero, M. I. | López-Bote, C.
A procedure to quantify intramuscular fat was developed using common inexpensive laboratory equipment. Three homogenization methods of lyophilized muscle samples (Ball-mill, Grinder and Mortar) and two extraction methods (Ball-mill or Vortex) were used in turkey meat and pork. Two-hundred mg of lyophilized and homogenized samples were accurately weighed and mixed with 1.5. mL of dichloromethane-methanol (82) and shaken either in a Mixer Mill (MM400, Retsch Technology) or in a Vortex. The final mixture was separated by centrifugation. Solvent was evaporated under a nitrogen stream and lipid content was gravimetrically determined. Besides, it was checked that the fatty acid profile was not altered by the protocol used. Moreover, the analysis of 4 replicas from the same sample showed different variation coefficients (16-29%) for the new procedures proposed over a wide range of IMF content. The combination of Grinder and Vortex methodologies can be proposed as a simple and inexpensive alternative to previous ones. © 2015.