Evaluation of oscillometric and vascular access port arterial blood pressure measurement techniques versus implanted telemetry in anesthetized cats
2011
Zwijnenberg, Raphael J. | Rio, Carlos L del | Cobb, Rami M. | Ueyama, Yukie | Muir, William W.
Objective—To compare the use of a semi-invasive vascular access port (VAP) device or noninvasive oscillometry versus invasive telemetry for blood pressure measurements in cats. Animals—6 healthy cats. Procedures—30 days before the study, all cats received an implanted telemeter and a VAP device. During normotension and experimentally induced hypertension, blood pressure was measured with the implanted devices and with noninvasive oscillometry at 4 time points. Results—Compared with invasive telemetry, VAP had a correlation coefficient from 0.8487 to 0.9972, and noninvasive oscillometry had a correlation coefficient from 0.7478 to 0.9689. Conclusions and Clinical Relevance—Use of the VAP device and noninvasive oscillometry had a high degree of correlation with invasive telemetry as the gold standard for blood pressure measurement. Use of a VAP device resulted in a slightly higher degree of correlation, compared with noninvasive oscillometry.