Refinar búsqueda
Resultados 1-2 de 2
Evaluation of isoflurane and sevoflurane vaporizers over a wide range of oxygen flow rates
2006
Ambrisko, T.D. | Klide, A.M.
Objective-To examine the accuracy and precision of isoflurane and sevoflurane anesthetic vaporizers. Sample Population-5 identcal isoflurane and 5 identical sevoflurane vaporizers. Procedures-Oxygen flow rates from 0.02 to 10 L/min were used with different vaporizer dial settings. Agent concentrations were measured at the common gas outlet by use of a refractometer. Accuracy was defined as the difference between measured agent concentrations, and dial settings were expressed as a percentage of the applied dial settings. Precision was defined as SD of the measured agent concentrations for each combination of dial setting and flow. Results-Isoflurane values were generally greater than the dial settings. Accuracy of the isoflurane vaporizer was > 20% when 0.6% and 1% was dialed. Accuracy of the sevoflurane vaporizer was always within +/- 20% but decreased at 0.02 L/min flow and at combinations of high flow and high dial settings. Overall precision of the isoflurane vaporizer was better than that of the sevoflurane vaporizer. Conclusions and Clinical Relevance-A possible explanation for the inaccuracy of the isoflurane vaporizer may be that it was manufactured to be accurate with air but not oxygen, which must be accounted for when using the vaporizer with oxygen, especially with nonrebreathing systems. The sevoflurane vaporizer may not deliver accurate agent concentrations at high flow and high dial settings. Both vaporizers are suitable for clinical use with a wide range of oxygen flow rates if these precautions are properly addressed.
Mostrar más [+] Menos [-]In vitro mechanical evaluation of torsional loading in simulated canine tibiae for a novel hourglass-shaped interlocking nail with a self-tapping tapered locking design
2006
Dejardin, L.M. | Lansdowne, J.L. | Sinnott, M.T. | Sidebotham, C.G. | Haut, R.C.
Objective-To describe a novel interlocking nail (ILN) and locking system and compare the torsional properties of constructs implanted with the novel ILN or a standard 8-mm ILN (ILN8) by use of a gap-fracture model. Sample Population-8 synthetic specimens modeled from canine tibiae. Procedures-An hourglass-shaped ILN featuring a tapered locking mechanism was designed. A synthetic bone model was custom-made to represent canine tibiae with a 50-mm comminuted diaphyseal fracture. Specimens were repaired by use of a novel ILN or an ILN8 with screws. Specimens were loaded for torsional measurements. Construct compliance and angular deformation were compared. Results-Compliance of the ILN8 was significantly smaller than that of the novel ILN. Mean +/- SD maximum angular deformation of the ILN8 construct (23.12 +/- 0.65 degrees) was significantly greater, compared with that of the novel ILN construct (9.45 +/- 0.22 degrees). Mean construct slack for the ILN8 group was 15.15 +/- 0.63 degrees, whereas no slack was detected for the novel ILN construct. Mean angular deformation for the ILN8 construct once slack was overcome was significantly less, compared with that of the novel ILN construct. Conclusions and Clinical Relevance-Analysis of results of this study suggests that engineering of the locking mechanism enabled the novel hourglass-shaped ILN system to eliminate torsional instability associated with the use of current ILNs. Considering the potential deleterious effect of torsional deformation on bone healing, the novel ILN may represent a biomechanically more effective fixation method, compared with current ILNs, for the treatment of comminuted diaphyseal fractures.
Mostrar más [+] Menos [-]